Controls and Connectors
Monitor Installation and Patient Safety 11-31
Monitor Installation
and Patient Safety
Caution
The System Power Outlet cannot be used to output power to other
devices such as the Philips M1167A/77A ACMS.
The System Power Outlet can only be used to supply power to the
ACMS when the total leakage current does not exceed 300 µA.
4. Equipotential Grounding Terminal; this is used to connect the
Anesthetic Gas Module to the hospital´s grounding system.
5. Line protection fuses, T1.6 H 250V.
6. Anesthetic Gas Exhaust.
Caution
Make sure that the anesthetic gas exhaust at the rear of the module is
connected to the gas extraction system. If it is not, contact your
biomedical engineer in order to make this connection.
The module can be operated from an ac power source of 100 - 240 V
±
10%, 50/60Hz.
The adjustment is made automatically by the power supply inside the
module.
Assembling
the System
The Systems consist of the following separate parts:
The Display Module (with control panel)
The Parameter Modules (with parameter racks)
The Computer Module
The Anesthetic Gas Module (optional)
If the system (display module and computer module) was delivered as an
unpackaged shipment, the assembly procedure below does not apply.